In 2000, Comfortex designed, tested and introduced the Landing Strip floor cushion. It proved to have an ability to minimize impact forces to protect patients from bedside injury should a fall occur. Thousands of Landing Strips, now serve residents and healthcare providers across the nation.
Recent testing done by the TUV/NTS, Detroit Division, further illustrates the designs superiority over much thicker and more cumbersome pads.
The Testing Laboratory dropped a 16 lb. weight from heights of 12’’, 18” and 24”. Testing measured G-forces generated upon impact and the products ability to de-accelerate the falling object gradually vs. suddenly.
(example : a falling object attached to a bungy cord de-accelerates over a greater time span vs. an object suddenly stopped by a non-flexible cable).
| Impact Test Results |
(peak G-Fores) |
|
| |
12” |
18” |
24” |
Avg. De-Acceleration Time |
| 1” Landing Strip |
57 |
101 |
152 |
8.0 mili. sec. |
2.5” 30 IFD Foam Pad
Vinyl Covered |
211 |
630 |
740 |
1.6 mili. sec. |
2.5” 30 IFD Foam Pad
Nylon Covered |
427 |
656 |
858 |
1.0 mili. sec. |
| Solid Floor |
726 |
1014 |
1390 |
0.4 mili. sec. |
- * Impact reductions were 4 to 9 times better with the Landing Strip design.
- * De-Acceleration was 5 to 20 times better with the Landing Strip design.
- Test Report available upon request